Electronic caliper L 150 0,01 IP54

Artykuły Techniczne Artykuły Techniczne
0000005402
Measuring range: 150 mm8 sizes of this caliper
€40.46 tax inc.

Net price: €32.89

Instant Shipping
W magazynie: 10-100 szt.
W magazynie: 10-100 szt.

Instant Shipping

Call and order: (44) 755-04-05

Description

Electronic caliper 150 mm IP54 – technical data

Instrument range, resolution, accuracy, and protection rating

Technical data
Measuring range 0–150 mm
Resolution 0.01 mm
Accuracy ±0.02 mm according to DIN 862
Protection rating IP54 according to IEC 60529
Reading LCD display
Units mm / inch
Measurement types external, internal, depth, step
Material hardened stainless steel
150 mm IP54 electronic caliper — product drawing with LCD and a reading of 17.00 mm
0–150 mm
Measuring range
0.01 mm
Resolution
±0.02 mm
DIN 862 accuracy
IP54
Protection rating

Electronic caliper 0–150 mm replaces four instruments: it measures diameters, holes, depths, and steps. You read the result in 0.01 mm increments directly from the LCD — without counting vernier marks or parallax error. The IP54-rated housing protects the electronics against dust and water splashes.

The beam and jaws are made of hardened stainless steel. The ORIGIN button activates absolute measurement: the caliper remembers its reference point after being switched off and is ready immediately, without zeroing. The ZERO button zeros the reading at any slider position for comparative measurements, while mm/in switches the result to inches.

Digital caliper construction – components and their functions

One jaw is fixed, while the other moves together with the display

Electronic caliper construction: jaws, beam with scale, locking screw, display, and extending depth rod
Drawing based on the product photograph. The jaws are open to 17.00 mm — exactly the value shown on the display. Proportions are illustrative.

Move the sliding jaw up to the workpiece, and the electronics convert the slider position into a dimension. The locking screw maintains the opening, and the thumb roller helps provide smooth adjustment — however, you control the measuring pressure yourself, because the caliper does not have a micrometer ratchet.

Four types of measurement with one instrument

External, internal, depth, and step — the same readout

External dimension
The lower jaws fit around a shaft, flat bar, or screw head. Measure as close to the beam as possible to reduce tilt error.
Close the jaws gently: excessive pressure bends the arms and produces a low reading.
Internal dimension
Spread the upper knife-edge jaws inside a hole or groove until they contact the walls.
Position the jaws across the centerline of the hole and use the largest reading.
Measuring depth with a caliper
Rest the end face of the beam flat on the edge of the hole and extend the depth rod to the bottom.
Keep the beam perpendicular — tilting produces a high reading. The depth rod must touch the bottom, not a chip.
Step height
Place the end face of the beam on the higher surface and rest the depth rod on the lower surface. The screen will show the difference in height.
Press the entire end face against the surface without rocking the instrument.
Comparing a series of workpieces with the ZERO function

Measure a reference, for example 25.00 mm, and press ZERO. Subsequent workpieces will be displayed as a deviation, for example +0.37 mm, without mental subtraction. After the series, close the jaws and zero the instrument again.

Absolute measurement — ORIGIN button

Close the jaws until they touch and press ORIGIN — the display will show 0.00 mm, and the caliper calculates absolute dimensions from this reference. The reference point is retained after power-off and automatic sleep: switch the instrument on and measure immediately. Set ORIGIN once — reset it only after replacing the battery.

A closer look at IP54 – protection against dust and splashing water

The IP code according to IEC 60529, read digit by digit

Read the IP code from IEC 60529 digit by digit: the first refers to solid objects, the second to water. An IP54 digital caliper operates where grinding dust and oil mist are present.

IP 5 4
Digit 5 — solid particles
Dust does not enter in quantities that interfere with the electronics
level 5 of 6
Digit 4 — water
Splashes from any direction do not harm the instrument
level 4 of 9
IP54THIS PRODUCT
Dust and splashes — workshop, production floor, measuring room
IP65
Dust-tight protection and water jets — parts washing area
IP67
Immersion 1 m / 30 min — coolant, wet machining, CNC
Why moisture affects a capacitive encoder

The dielectric permittivity of water is approximately 80 times that of air. A film of moisture in the gap changes capacitance locally and distorts the reading before the electronics are damaged. Wipe the beam and guide dry after use.

A drop of water in the caliper encoder gap disturbs the electric field between the electrodes
Limitations of the IP54 rating

Do not immerse the instrument in coolant or rinse it with a water jet. The IP code does not confirm resistance to oils or solvents. For wet machining, choose an IP65 or IP67 caliper.

Accuracy of ±0.02 mm according to DIN 862 – resolution is not error

Maximum permissible errors (MPE) and the effect of temperature on the result

A resolution of 0.01 mm only indicates the smallest display increment. Reliability depends on accuracy: DIN 862 permits an error of ±0.02 mm for external measurements and ±0.04 mm for internal or depth measurements over the 0–150 mm range.

Measurement typeMaximum permissible error (MPE) according to DIN 862
External (lower jaws) ±0.02 mm
Internal (knife-edge jaws) ±0.04 mm
Depth (depth rod) ±0.04 mm
Workshop example — effect of temperature

Steel expands by approximately 11.5 µm per meter per degree Celsius. A 150 mm workpiece that is 10°C warmer than the instrument will read 11.5 × 0.15 × 10 ≈ 17 µm larger — almost the entire permissible error. Measure at a temperature close to 20°C.

IP54 electronic caliper handles the full range of everyday measurements up to 150 mm, while the digital readout speeds up batch inspection and eliminates errors when transcribing results. When tolerances fall below 0.05 mm, use a micrometer — a caliper does not comply with Abbe's principle and cannot achieve its accuracy.

DID YOU KNOW? How a capacitive encoder works – from movement to digits

The electronics do not read scale marks. The slider board with powered pads and the strip with passive reflectors form a capacitor with variable electrode overlap — the signal crosses the air gap, and the two boards never touch. A T-shaped reflector reflects the signal from the transmitting pads to the receiving pad. Moving the slider by one pattern period shifts the signal phase by 360°; the circuit divides the approximately 1 mm period into 100 steps — hence 0.01 mm on the display.

Cross section of a caliper capacitive encoder: transmitting pads, air gap, and T-shaped reflectors
Simplified sensor cross section with an enlarged gap. Colors distinguish electrode functions, not actual materials.

DID YOU KNOW? How a capacitive encoder works – from movement to digits

The electronics do not read scale marks. The slider board with powered pads and the strip with passive T-shaped reflectors form a capacitor with variable electrode overlap — the signal crosses the air gap and returns to the receiving pad. Moving the slider by one pattern period shifts the signal phase by 360°; dividing the approximately 1 mm period into 100 steps gives 0.01 mm on the display.

Caliper capacitive encoder diagram: electric field between the pads and T-shaped reflectors
Dashed lines: red — electric field in the gap, green — signal path to the receiving pad.

IP54 caliper – frequently asked questions

It is a code from IEC 60529: digit 5 indicates protection against dust in quantities that interfere with operation (level 5 of 6), while digit 4 indicates resistance to water splashes from any direction (level 4 of 9). The instrument is suitable for workshops and production floors, but not for use in a coolant jet or for immersion.

Resolution is the smallest display increment. Accuracy is the agreement of the result with the true value — according to DIN 862, it is ±0.02 mm for external measurements over the 0–150 mm range. A reading of 25.37 mm therefore indicates a true value of approximately 25.35–25.39 mm.

Most often, ZERO was pressed with the jaws open, or dust or a chip remains between the surfaces. Clean the jaws, close them gently, and zero the instrument. If the deviation returns each time the instrument is switched on, reset the reference point using ORIGIN with the jaws closed. If the reading is still off, check the caliper against a gauge block.

Keep the guide and jaws clean, wipe the instrument dry after use, and check zero with the jaws closed. Calibrate the caliper with gauge blocks every 6–12 months. For tolerances below 0.05 mm, use a micrometer.

Reviews

No customer reviews for the moment.

Add a review about the product

Electronic caliper L 150 0,01 IP54

Electronic caliper L 150 0,01 IP54

Shipping

All orders are delivered by courier company GLS. For orders exceeding 369 PLN gross shipping is FREE

Courier GLS Przelew24 / Prepayment 14,76 zł gross Delivery time 1 working day
Courier GLS Cash on delivery 17,26 zł gross Lead time 1 working day

This website uses its own and third-party cookies to improve our services and show you advertising related to your preferences by analyzing your browsing habits. To give your consent to its use, press the Accept button.

More information